Maintenance • 61
9. MAINTENANCE
9.1 Information about maintenance
Your seca scales leave the factory with an accuracy
greater than ±0,15%. To preserve this level of accuracy
the product must be set up carefully and serviced regu-
larly. We recommend having it serviced every 3 to 5
years depending on how often the scales are used.
... only the item "SYS" is
visible in the rf menu?
• The wireless module is disabled.
- Activate wireless module (see "Activate wire-
less module (SYS)" on page 56).
... only the items "SYS"
and "lrn" are visible in
the rf menu?
• The wireless module is activated but no
wireless group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 56).
... the items "APrt" and
"Time" are not visible in
the rf menu?
• No wireless printer is logged into the wireless
group.
- Use menu item "lrn" to log the wireless printer
into the wireless group(see "Set up wireless
group (Lrn)" on page 56).
…
Er:X:11 is displayed?
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight-
more evenly
- Restart scales
…
Er:X:12 is displayed?
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
…
Er:X:16 is displayed?
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales
… when the Enter key
(send/print) is pressed
and
Er:X:71 is dis-
played?
Data transmission not possible, wireless module
is disabled.
- Activate wireless module (see "Activate wire-
less module (SYS)" on page 56).
… when the Enter key
(send/print) is pressed
and
Er:X:72 is dis-
played?
Data transmission not possible, no wireless group
set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 56).