EasyManua.ls Logo

Seca 704 - Weighing Data

Seca 704
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66 •
11.2 Weighing data
Power consumption
with wireless module disabled
with wireless module activated
approx. 32 mA
approx. 50 mA
Maximum runtime in battery mode
with wireless module disabled
with wireless module activated
approx. 5.600 minutes
Power supply unit recom-
mended
Medical device according to directives 93/42/
EEC and 2007/47/EC
Class I with measuring function
Calibration according to directive 2009/23/EC Class III
EN 60 601-1:
Class II totally insulated appliance:
Type B electromedical device:
Radio transmission
Frequency band
Transmission power
Observed standards
2,433 GHz - 2,480 GHz
< 10 mW
EN 30028, EN 301 489-1, -17
Technical data for
seca 704
Weighing data
Calibration according to directive 2009/23/EC Class III
Maximum load
Weighing range 1
Weighing range 2
150 kg
300 kg
Minimum load
Weighing range 1
Weighing range 2
1 kg
2 kg
Graduations
Weighing range 1
Weighing range 2
50 g
100 g
Tare range up to 300 kg
Accuracy following initial calibration
Weighing range 1: 0 to 25 kg
Weighing range 1: 25 to 100 kg
Weighing range 1: 100 kg to 150 kg
Weighing range 2: 0 to 50 kg
Weighing range 2: 50 to 200 kg
Weighing range 2: 200 kg to 300 kg
± 25 g
± 50 g
± 75 g
± 50 g
± 100 g
± 150 g

Table of Contents

Other manuals for Seca 704

Related product manuals