Figure5‐8Measurementdisplay
5) Ifnecessary,youcanpressthe"Save/▼"keytoSavethemeasurementresults.
Toensure accurate measurement, the sample volumeinthecontainershouldbe sufficient toensure that the
junctionofthemeasuringandthereferenceelectrodeiscompletelyimmersedinthesample.
Foraccuratemeasurement,itisrecommendedthatusers calibrate
and
measureat thesametemperature.
6
MaintenanceandTroubleshooting
6.1 Maintenance
Thecorrectuseandmaintenanceofthemetercanensurethenormalandreliableoperationofthemeter,
especiallythepHmeter,whichhasahighinputimpedanceandiseasilydamagedbystaticelectricityandother
electromagneticinterference;Inaddition,itwilloftencomeintocontactwithchemicals,andtheuseenvironmentis
relativelyharsh,soreasonablemaintenanceisrequired:WhentheQ9interface ofthemeter is notconnectedto the
electrode,pleaseinserttheQ9short‐circuitplugtopreventthemeterfrombeingdamaged.
PleaseplacetheQ9short‐circuitpluginadryandcleanenvironmenttopreventtheshort‐circuitplugfrombeingcorroded
toaffecttheshort‐circuiteffect.
•
Ifthemeterisnotusedforalongtime,pleasedisconnectthepowersupply.
•
Theelectrodesocketofthemetermustbekeptcleananddry,andshouldnotbeincontactwithacid,alkaliorsalt
solution.
•
Thehousingmaterial ofthismeterissensitivetocertainorganicsolventssuchastoluene, xylene,andmethylethyl
ketone(MEK).Ifliquidentersthehousing,itmaydamagethemeter.Tocleanthemeterhousing,wipegentlywitha
toweldampenedwithwaterandmilddetergent.
•
Ifthemeterisnotusedforalongtime,pleasecoverthedustcover.
•
Whentransportingthemeter,pleaseabidebythefollowinginstructions:
•
Unplugthemeterandremoveallconnectedcables.
•
Removeelectrodeholder.
•
Toavoidmeterdamageduringlongdistancetransportation,pleasetrytouseoriginalpacking.
6.2 Electrodeusageandmaintenance
Before using the electrode, you should read the electrode manual carefully to know the type, structure and
applicationscopeoftheelectrode.ForcompositepHelectrodeswithplasticshells,thefollowingpointsshouldgenerally
benoted:
The external reference supplement of the composite electrode is 3mol/L potassium chloride solution. The
supplementcanbeaddedfromthesmallholeatthetopoftheelectrode.Whenthecompositeelectrodeisnotin
use,therubbercovershouldbepulledtopreventthesupplementfromdryingup.