Questions and Answers

  • C
    Curtis EvansAug 12, 2025
    Why doesn't the Sentry Recording Equipment recognize the end-of-batch signal?
    • M
      Monique FryAug 12, 2025
      First, ensure the Sequencer is in Batch mode. If not, switch to Batch mode. If it is already in Batch mode, verify that the End-of-Batch indication is correctly wired between the analyzer and Sequencer.
  • A
    Andrew HawkinsAug 21, 2025
    What to do if the Sentry Sample Sequencer 5 Recording Equipment display doesn’t light up?
    • M
      mmcmahonAug 21, 2025
      First, check the power LED on the circuit board. If it's not lit, check the power connections from the Power Supply to the circuit board to ensure they are tight. If the connections are secure, check the Power Supply fuse for continuity; replace the fuse if it's bad.
  • J
    Joseph FrySep 1, 2025
    How to fix incorrect analyzer value on Sentry Sample Sequencer 5 Recording Equipment?
    • M
      marymarshSep 1, 2025
      First, verify the scaled output of the analyzer matches the scaled input of the Sequencer. If the issue continues, perform an analog input calibration.