can no longer be used for an alarm signal.
6. After pressing ENTER, return the #1 DIP switch (S1) to the OFF position. Replace the
cover and close the meter panel. This returns the analyzer to normal operation. Note: If a
"cold start" is initiated or the backup battery fails during a power failure, you must
reset the MARK END of MEAS as outlined above.
7. Verify the Sample Sequencer is operating in the batch mode.
Recommended Sequencer Timer Settings
It is recommended the following times be programmed into the Sample Sequencer when using
the Hach Series 5000 analyzers:
Sample Time.......... 20:00 minutes
Reference the Operation section of this manual for instructions on programming the Sample
Timers.