EasyManua.ls Logo

serva Blue Series - 4.5 Start the run; 4.6 After the run

serva Blue Series
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
4.5 Start the run
1) Close the lid while lowering the electrodes on the wicks. Plug in the leads.
2) Switch the thermostatic circulator on, set to
HPE Gels: 15 °C
SDS Gels: 15 °C
Clean Gels: dependent on the application (see gel manual)
PreCotes, PreNets: 10°C
CleanGel IEF: 10°C
Focus Gel: dependent on the application (see gel manual)
Note: During Electrophoresis the electric resistance of the gel is slowly increasing.
Therefore, the heat production during the starting phase is rather low. It does not
cause overheating when the chiller begins cooling at the same time the
electrophoresis is started.
3) Start the run according to the settings described in the gel manual.
DOMINIQUE DUTSCHER SAS