1 Safety
The SERVA HPE
TM
BlueHorizon comply with the standards and directives mentioned
in the applicable CE declaration.
Please take note of the following safety measures:
Warning: Operation of this instrument requires high voltage.
Disconnect the high voltage external power supply before opening
any drawer.
Turn off and disconnect any high voltage power supply before
opening the safety lid.
Disconnect the high voltage external power supply and the AC
main supply before cleaning or servicing.
Do not spill or store liquids on top of the unit.
If liquid is spilled into the HPE
TM
BlueHorizon, disconnect the high
voltage power supply and the AC main power immediately before
opening the safety enclosure lid.
Do not operate or connect power sources to the equipment if
there is any mechanical damage.
The supplied DC cables are rated for 5,000V. Only use cables
and adaptors supplied with the HPE
TM
BlueHorizon or ensure that
these have a suitable DC insulation compliance for the used
voltages.