7.1
SECTION 7 TECHNICAL SPECIFICATIONS
7.1 Introduction
This section includes the technical specifications for all versions of the 4000. The user must
ensure that the relevant sub-sections are used for reference.
It may be noted that similar transducer options are available in different analyser variants, in
some cases the specifications for these will be application dependent.
(This performance specification has been written, and verified, in accordance with the
international standard IEC 1207-1:1994 "Expression of performance of gas analysers").
7.2 Generic 4000 series analyser performance
7.2.1 Environmental specifications
Operating temperature: 4100D 5 to 40°C / 41 to 104°F
4200D/4210D 5 to 40°C / 41 to 104°F
4900D 5 to 45°C / 41 to 113°F
Storage temperature: All analysers -20 to 60C / -4 to 140F
Relative humidity: 10 to 90% HR, non-condensing.
Atmospheric pressure: 79 to 124kPaa / 11 to 18psia
(for operating altitudes to 2000m)
Installation category: II (local level power distribution with over voltage to
withstand up to 2500Volts) in accordance with IEC 664
Pollution degree 2 (normally electrically non-conducting pollutants) in
accordance with IEC 664
Warm up time: Typically 24 hours from cold start at 20°C/68°F, may be
longer for some high sensitivty measurements.
7.2.2 Power supply
The analyser is CSA/FM Approved for only 110-120 / 220-240Vac (+/-10%),50/60Hz, 350VA
maximum, but will function correctly for 85-132/170-264Vac,47-62Hz, 350VA maximum.
7.2.3 Design standards
The analyser conforms to the following normalised European standards for performance,
product safety and electromagnetic compatibility:
EN61010-1
Safety requirements for electrical equipment for measurement, control and laboratory
use.