EasyManua.ls Logo

Sharp SX68JF100 - Page 7

Sharp SX68JF100
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
SX68JF100
7-1 7-2
adjust
Fig. 5( a)
Fig. 5 (b)
Fig. 6 (a)
OFF SET
Fig. 7 (a)
Fig. 6 (b)
Fig. 7 (b)
OFF SET
NO. Adjustment part Adjusting procedure and conditions Waveform and others
SECAM CHROMA ADJUSTMENT
1 BELL FILTER f0
(SCM BELL)
I
2
C bus
adjustment
1 Receive Oi-5CH(SECAM colour bar).
2 Select the "SCM BELL" fo adjustment mode.
3 Connect the oscilloscope to TP901 of main.
(TP2901 in VCJ unit)
Range : 10mV/Div. (AC)
Sweep time: 20µsec/Div. (10:1 probe is used.)
4 Vary the bell filter data to align the waveform
to be the nearest to that in Fig. 5 (b).
(Arrange the level to gain the whole balance.)
2 SECAM black
level R-Y/B-Y
(R-Y and B-Y)
I
2
C bus
adjustment
1 Receive Oi-5CH(SECAM colour bar).
2 Select SECAM black level adjustment "R-Y"
mode.
3 Connect the oscilloscope to TP802 (TP2804
in VCJ unit) of the main.
Range: 10mV/Div.
Sweep time: 20µsec/Div. (10:1 probe is used.)
4 Vary R-Y data to minimize the offset between
the non-signal line and signal line as shown in
Fig. 6 (b).
5 Select SECAM black label adjustment "B-Y"
mode.
6 Reconnect the oscilloscope to TP803 (TP2805
in VCJ unit) of the main.
The conditions are the same as in Item (3).
7 Vary B-Y data to minimize the offset between
the non-signal line and signal line as shown in
Fig. 7 (b).
* To be at the same levbel.
NO. Adjustment part Adjusting procedure and conditions Waveform and others
1 SUB SOUND
(SUB VOL)
(I
2
C bus
adjustment)
1 Receive E-12ch (PAL colour bar). Signal con-
tent: 400Hz 100% Mod.
2 Connect the probe of the meter to (SI) con-
nector. (See Note)
3 Select "SUB VOL" in the service mode.
4 Adjust the sub-sound data to make the meter
indicate 8.5Vrms (at L-ch).
Adjustment value : 8.5 ± 0.4Vrms
NO. Adjustment part Adjusting procedure and conditions Waveform and others
SIF (NICAM/IGR) ADJUSTMENT
1 VCO COIL
T2300
(in the NICAM
unit)
1 Receive E-69CH (PAL colour bar).
(Set the receiving frequency at AFT OFF Ex.
E-69=855.25 MHz.)
2 Connect the digital voltmeter to TP301 (in the
main unit).
3 Verify that it is turned counterclockwise to 0V
and clockwise to 5V, and adjust T2300 to make
the DC voltage of TP2300 become 2.5 ± 0.1V
in the range.
<Single-unit adjustment> - For your Reference.
Vcc 5 ± 0.1V
IF input frequency 38.9MHz ± 10kHz
Control it under these conditions.
Adjust T2300 to make the DC voltage of TP2300
(in the NICAM unit) become 2.5 ± 0.1V.
After single-unit adjustment, check the practical
setup.
Align the receiving frequency to the select chan-
nel frequency. (AFT OFF)
The checked TP301 in the main unit voltage must
be 2.5 ± 1.0V.
(Take care that ±1.0 is considered for the Vcc dif-
ference between the single unit and unit, Fo dif-
ference and so on, but is not the adjustment pre-
cision.)
2 Noise mute
check
1 Receive E-12CH (PAL colour bar).
2 Maximize the sound volume to verify that the
sound is output from the speaker. Then, set
the non-signal state.
3 At this time, verify that the audio mute oper-
ates.
4 After checking the operation, set the sound
volume to be the minimum.
Note:
* SPATIALIZER ..... OFF
* S-Normal state
* S-VOL ................ Max
* S-BOOSTER ...... OFF
SOUND LEVEL ADJUSTMENT

Related product manuals