World first! –– flame extinguished automatically by vibration sensor
The flame is automatically extinguished when the new built-in sensor detects vibrations.
Instruments are completely safe during severe shaking caused by earthquakes.
Multimode automatic gas leak check
The fuel gas pipes inside the instrument gas controller are automatically
checked for gas leaks when the power is turned on and when the flame is
extinguished. Should a gas leak be detected, a warning buzzer sounds and a
warning is displayed on the screen.
Flame-retardant materials
UL 94V-0-rated flame-retardant materials
are used for the instrument external
covers and atomizer unit. This superior
design pays due consideration to safety.
Safe ignition switches
The switches perfectly fit finger contours and are positioned
near the window.
To prevent inadvertent operation, both switches must be
pressed simultaneously to start ignition.
Durable pipes and reliable couplings
All pipes are selected for durability and
used with highly reliable couplings.
Clear window
The large window permits observation of the flame, and the
burner can be easily accessed through the large opening.
The window closes when released to eliminate the chance
of it being left open.
Automatic flame ignition and extinguishing
Flame ignition and extinguishing operations are extremely simple. The
Air-C
2
H
2
flame priority ignition mechanism prevents flashback.
Automatic Air-N
2
O switching system with acetylene flow-rate monitor
After the ignition of an Air-C
2
H
2
flame, the flame automatically switches to
an N
2
O-C
2
H
2
flame.
If the C
2
H
2
flow rate does not increase due to solenoid valve trouble, for
example, switching to the support gas is disabled to prevent flashback.
Pressure monitor prevents flashback Automatic gas shut-off by flame monitor
Sudden power interruption detection and re-ignition safety measures
Drain tank level monitor
Mechanism to prevent misuse of the burner
Safety Functions for Flame-Specification Instruments
Safety Functions for Furnace-Specification Instruments
The connected options and operation of safety mechanisms are
checked during instrument initialization.
Cooling water flow rate monitor Argon gas pressure monitor Furnace cooling check
Overcurrent protection unit (double-checked by circuit protector and optical sensor)
Safety-Conscious Design