ADM-860C 04/14/0815
STORE FULL
This message indicates that the number of readings in memory has reached the maximum storage capacity
of 100 readings.
STORE MODE
This message is displayed when a key other than READ, STORE or RCL is pressed when readings are in
memory. This message may be displayed when the SHIFT key is pressed to indicate that the meter is in the
STORE mode. The mode may not be changed until the readings in memory are cleared by pressing
SHIFT/CLEAR.
STORE RDY
This message indicates that the meter has been placed in the memory mode, and that readings will be
sequentially stored in memory. This function permits recall of readings using the RCL keys, or of their sum,
average, minimum or maximum by pressing the STORE key at any point.
TEMP
This signal will be flashed when the Temperature mode is selected, and also upon each subsequent
operation of the READ key.
TEMPROBE
This message is displayed to indicate a change in the TemProbe connection status during a STORE
sequence.
TOO HOT or TOO COLD
If the internal temperature of the meter exceeds its operational limits, the meter will display TOO HOT or
TOO COLD and shut down. However, if there are readings in memory, the meter will continue to display
TOO HOT or TOO COLD, and will retain the readings in memory. The meter must be cooled down or
warmed up, as the case may be, before normal operation can resume.
If the meter has displayed OVER RANGE after displaying either TOO HOT or TOO COLD, but has not shut
down, this message indicates that the TemProbe sensor was being exposed to temperature levels beyond
the proper operating range. If TOO HOT/OVER RANGE has been displayed, but the meter has not shut
down, the TemProbe sensor may be short circuited.
TOO LOW
This message indicates that the battery charge is very low and the back-light may not be used until the
batteries have been recharged (NO LIGHT/BATTERY/TOO LOW).
TREND RDY
This message is displayed when the meter has been placed in the TREND mode. TREND readings are not
certified for accuracy.
UNDER TEMP
This display advises the operator that the temperature measurement being attempted exceeds the lower
range of the meter.
UNITS - XX
This message will be displayed to indicate the units in which readings are being stored. It is also one of the
parameters displayed after SHIFT/SHIFT is pressed.
VELGRID
This signal will be flashed when the VelGrid mode is selected, and also upon each subsequent operation of
the READ key.