ADM-860C 04/14/0820
Press the STORE key, then the MODE key. PRINT MEM? will be displayed. Press the READ key to download
all readings stored in memory to the printer. PRINTING will be displayed while the readings are being printed.
The readings will remain stored in the meter memory until the memory is cleared by pressing the SHIFT key,
followed by the CLEAR key. An example of readings printed from memory in the VelGrid mode is shown below:
VELGRID
LOCAL DENS
01 1576 FP
02 1581 FP
03 1578 FP
04 1577 FP
Sum 6312 Fp
004 Avg 1578 F
001 Min 1576
002 Max 1581
Std Dev 2.2
An example of readings printed from memory in the FlowHood mode is shown below:
FLO-HOOD
LOCAL DENS
01 1474 CFu
02 1524 CFc
03 1482 CFu
04 1531 CFc
Sum c 3055
Sum u 2956
002 Avg c 1527
002 Avg u 1478
002 Min c 1524
001 Min u 1474
004 Max c 1531
003 Max u 1482
Std Dev c 4.9
Std Dev u 5.7
5.5 DATA DOWNLOAD TO A COMPUTER
Readings from the ADM-860C may be automatically transmitted to specific cells of a computer spreadsheet
such as Microsoft Excel™ for display, analysis and manipulation. This functionality requires either
HyperTerminal Private Edition or WinWedge software from TAL Technologies, Inc. as well as an RS232 serial
cable. The procedure using HyperTerminal is supplied with the ADM-860C AirData Multimeter.
WinWedge
collects data from scales and balances, gauges, laboratory instruments, meters, sensors or any
other RS232 instrument or serial device and inputs the data directly into Microsoft Excel or any other Microsoft
Windows™ application program.
Configuration files must be set up to specify how WinWedge will interact with the AirData Multimeter (or with
any other device). A series of fully functional, example Configuration Files is provided along with the WinWedge
software to help the user become familiar with the program. These Configuration Files are extremely easy to
use and are provided so the most basic set-up will occur automatically and provide the user with an introduction
to the available features. The Configuration Files may be copied repeatedly and each copy may be modified
as required to suit the needs of the user.
An easy step-by-step set of instructions titled GET A QUICK START USING THE WINWEDGE
CONFIGURATION FILES, is also provided along with the sample Configuration Files.