ADM-860C 04/14/0838
NOTICE: The use of more than one MultiTemp and temperature cable extender set may reduce the meter
reading by a non-linear degree depending on the combination of cable type, length and TemProbe temperature.
This is due to the added resistance of the additional MultiTemp(s) and cable(s). This effect is likely to be
negligible at very low temperatures, but may be
$ 0.25/ F per additional MultiTemp and cable extender at 154/
F. The offset correction(s) must be determined by comparing readings taken with and without the additional
MultiTemps and extender cables at the temperature(s) to be measured.