ADM-870C 07/14/0920
5.6.2 DOWNLOAD AUTOMATIC READINGS TO A PRINTER
Turn the printer on and then turn the meter on. The mode and units saved by the meter during the most recent use will be
printed. Select the desired mode and units and then press SHIFT/AUTO to place the meter in the automatic reading mode.
Press the READ key. The printer will print the mode and density selections, followed by each of the readings in direct
sequence. Hold the READ key down at any time to halt the reading sequence and stop the printing process. Pressing the
READ key again will start the automatic reading and printing process again. An example of readings printed from the
automatic reading mode is shown below.
VELGRID
LOCAL DENS
FP 1576
FP 1589
HALT
VELGRID
FP 1561
HALT
FP 1590
FP 1561
5.6.3 DOWNLOAD READINGS IN MEMORY TO A PRINTER
Readings stored in memory may be downloaded from the meter to a printer after the sequence of readings has been
completed. Connect the printer to the meter using the RS232 cable as described in Section 5.6 DATA DOWNLOAD TO
A PRINTER. Turn the printer and meter on.
Press the STORE key, then the MODE key. PRINT MEM? will be displayed. Press the READ key to download all readings
stored in memory to the printer. PRINTING will be displayed while the readings are being printed. The readings will remain
stored in the meter memory until the memory is cleared by pressing the SHIFT key, followed by the CLEAR key. An
example of readings printed from memory in the VelGrid mode is shown below:
VELGRID
LOCAL DENS
STD 70
/ F
01 1576 FP
02 1581 FP
03 1578 FP
04 1577 FP
Sum 6312 Fp
004 Avg 1578 F
001 Min 1576
002 Max 1581
Std Dev 2.2
An example of readings printed from memory in the FlowHood mode is shown below:
FLO-HOOD
LOCAL DENS
01 1474 CFu
02 1524 CFc
03 1482 CFu
04 1531 CFc
002 Avg c 1527
002 Avg u 1478
Sum c 3055
Sum u 2956
002 Min c 1524
001 Min u 1474
004 Max c 1531
003 Max u 1482
Std Dev c 4.9
Std Dev u 5.7