INSTALLATION 3
3.6.3 Connecting the signal connections (as required)
3.6.3.1 I/O module (standard)
The standard version has a built-in I/O module. A second I/O module can be fitted (option).
▸ Position of signal connections, see “Electrical connections”, page 17.
▸ Function of signal connections → “I/O module” Addendum to Operating Instructions.
▸ The cables have to be approved for the relevant application.
– Only use sufficiently screened cables.
– The shielding braid has to end in the cable duct.
– Shorten the shielding braid accordingly.
Fig. 5: Shielding braid
3.6.3.2 Intrinsically safe signal connections (option)
see “Intrinsically safe signal connections (option)”, page 20.
3.6.4 Connecting the interfaces (as required)
To use an interface:
▸ Connect the interface cable to the corresponding interface inside the enclosure, see
“Electrical connections”, page 17.
Function of interfaces → Operating Instructions “GMS800 Series”.