8018429/1DWP/V3-0/2022-06| SICK A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | GMS840
Subject to change without notice
TECHNICAL DATA 8
For GMS842 (
C
CSA
US)
8.5 Power connection
Power supply unit version: GMS842 (
C
CSA
US)
and GMS840
Power supply unit version: GMS841 (ATEX/ IECEx)
Parameter Allowable value
Sample gas pressure in sample gas
line:
–500 +1000 hPa (–0.5 +1.0 bar)
Sample gas volume flow: Max. 100 dm
3
/hour
Maximum purge gas pressure in
enclosure:
30 hPa (30 mbar)
Sample gas connection positions, see “Dimensions, fitting position and
connections”, page 27
Sample gas connections function “GMS800 Series” Operating Instructions.
Other sample gas specifications (pressure, volume flow etc.) Addendum to
Operating Instructions for the Analyzer modules fitted.
WARNING: A wrong power supply unit voids the approval
Only replace the power supply unit with a power supply unit of the same type.
Approval
CSA
UL
US
ClassI Div2 Group A-D T4
Power voltage: 85 ... 264 V AC (automatic switchover)
Power frequency (AC): 47 63 Hz
Permissible over-voltages:
Transient over-voltages in the supply network
should not exceed over-voltage category II
according to IEC 60364-4-443
Power input: 50 300 VA (depending on equipment)
Secondary fuse: 10 A (exchangeable fusible cutout)
[1]
Required connecting cable
Conductor cross-section: 3 times 1.5 mm
2
(AGW 16)
Version:
Cable listed according to CSA / UL, oil resistance
and flame retardancy have to be observed
according to the requirements
[1] F1 on the “fuse board”. Spare part: “ET fuse F10A0“, Part No. 2062251.
Approval IEC/EN 60079-15 Ex II3G EX nA IIC (T4) Gc
Power voltage
85 ... 132 V AC
187 ... 264 V AC
Power frequency (AC): 47 63 Hz
Permissible over-voltages:
Transient over-voltages in the supply network
should not exceed over-voltage category II
according to IEC 60364-4-443
Power input: 50 300 VA (depending on equipment)
Secondary fuse: 10 A (exchangeable fusible cutout)
[1]
Required connection cable
Conductor cross-section: 3 times 1.5 mm
2
(AGW 16)
Version:
Oil resistance and flame retardancy have to be
observed according to the requirements
[1] F1 on the “fuse board”. Spare part: “ET fuse F10AH250V”.