12
8019969/V1-0/2016-06 | SICKADDENDUM TO OPERATING INSTRUCTIONS | S710 E
Subject to change without notice
2 ENGLISH
2.3 Change between “mg/m
3
” and “ppm”
Note: Component oxygen (O
2
) is always shown in the physical unit % by vol. and cannot be
changed.
Fig. 3: Operating and display elements
1 Press “ESC” as often as required until you reach the main menu.
2 Press “.” (decimal point) to extend the menu selection.
3 Press key “6” to go to the “settings” menu.
4 A warning is displayed: Read the warning and confirm with “Enter”.
5 Press key “2” to go to the “measurement” menu.
6 Press key “3” for “meas. value display”.
7 Press key “1” to go to the “meas. component” menu. Select the measuring component
for which the physical unit is to be changed. Confirm your selection with “ENTER” and
exit the menu with “ESC”.
8 Press key “4” to go to the submenu where you can switch between “mg/m³” and “ppm”.
ModularSystem
Function
Service
Alarm
789
456
123
0
Esc
Help
Enter
Main menu
1 measuring display
2 device status
3 control
4 calibration
5 maintenance signal
Enter digit
No messages
CO 12 mg/m3
standard functions
operation information
status messages
measured values (alternating)
Main menu
1 measuring display
2 device status
3 control
4 calibration
5 maintenance signal
6 settings
7 service
Enter digit
No messages
CO 12 mg/m3
“settings”