14 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 14 of 34
Medical Solutions
X 10. Release fluoro and close X-iris until the X-ray field is 43 mm. Tolerance: + 2 mm/ -3 mm.
¹ See attached centering cross.
NOTE
For units with air kerma display it is important for software calcula-
tion that the X-iris closed position is adjusted accurately.
11. Stop fluoro and click on <Go>.
¹ Status bar: Closed Position --> running --> success
12. Click on <Go>.
¹ Status bar: Init Off & Store Positions --> running --> success
¹ A window appears: "Calibration is finished..." --> confirm with <OK>.
¹ Main system restarts.
13. Click on “>” for the next page.
¹ The main system restarts and the circle mask switches back to on.
¹ The X-iris positions can be configured manually here.
14. Configure the "Zoom Procedure" on this page per the country regulations or the cus-
tomer’s preference.
15. Click on <Finish>.
16. Check the adjustment and repeat if necessary.
Init Off & Store Values

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals