8 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 8 of 34
Medical Solutions
2- 2Adjustments
Generator Adjustment 0
1. Register an emergency patient to allow radiation release.
2. Place 2.1 mm Cu on the radiation exit port of the single tank.
3. Click on <Go>.
¹ Action bar: “start fluoroscopy”
X 4. Start fluoro
¹ Action bar: “continue fluoro”
¹ Status bar: “kV- offset-learning”
¹ Status bar: “kV- learning done”
¹ Action bar: “stop fluoroscopy”
5. Stop fluoro
¹ Status bar: “generator learning”
¹ Action bar: “start fluoroscopy”
X 6. Start fluoro
¹ Action bar: “continue fluoro”
¹ Status bar: “warmup” (~ 8 minutes)
¹ Status bar: “filament correction”
¹ Status bar: “learning pushfactor”
¹ Status bar: “gen. learning done”
¹ Action bar: “stop fluoroscopy”
7. Stop fluoro
8. Click on <Save>.
¹ A window appears: “Main system values were successfully saved” --> click on
<OK>.
9. Click on “>” for the next page (multiple exposure) and perform a generator adjustment
check.