20 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 20 of 34
Medical Solutions
X 12. Briefly release fluoro for the LIH image.
13. Adjust the display lines on the monitor until the lines are aligned with the front edges of
the slot diaphragm blades.
14. Click on <Go>.
¹ Status bar: Open Pos. I.I. - Zoom Format --> running --> success
15. Click on <Go>.
¹ Status bar: Init Off & Store Position --> running --> success
¹ A window appears: “Calibration is finished...” --> confirm with <OK>.
¹ The main system restarts.
16. Click on “>” for the next page and click on <Finish>.
17. Check the adjustment and repeat if necessary.
Init Off & Store Position