EasyManuals Logo

Siemens ARCADIS Varic Manual

Siemens ARCADIS Varic
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
16 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 16 of 34
Medical Solutions
15. Click on <Go>.
¹ Status bar: Init Off & Store Values --> running --> success
¹ A window appears: "Calibration is finished..." --> confirm with <OK>.
¹ The main system restarts.
16. Click on “>” for the next page and click on <Finish>.
17. Check the adjustment and repeat if necessary.
Open Pos. I.I. Zoom Format

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors