16
HART add-on module 10. Device-Specific Commands
Command Number and Data in Command Data in reply
Function
171 Read totalizer Byte 0 Totalizer no. Byte 0 Unit code
0 = Totalizer 1 Byte 1-4 Totalizer value
1 = Totalizer 2 Byte 5 Totalizer no.
172 Read status data Byte 0 log type Byte 0 As in command
0 = Error Pending Byte 1-18 9 logs consisting
1 = Status Log of two bytes,
Canapp_ID + Error
Number
173 Read status timestamp Byte 0 log type Byte 0-2 As in command
Byte 1 Canapp_ID Byte 3-12 days, hours and
Byte 2 ErrorNumber minutes
180 Write Currentoutput1 Byte 0 0 = Off As command
(MASS 6000 only) 1 = Massflow
2 = Fraction A
3 = Fraction B
4 = Volumeflow
5 = Sensor temperature
6 = Density
7 = Pct. Fraction A
181 Read Currentoutput1 None Byte 0 0 = Off
(MASS 6000 only) 1 = Massflow
2 = Fraction A
3 = Fraction B
4 = Volumeflow
5 = Sensor
temperature
6 = Density
7 = Pct. Fraction A
182 Write Digitaloutput1 Byte 0 0 = Off As command
1 = Pulse
2 = Frequency
3 = Error level
4 = Error number
5 = Direction/Limit
6 = Batch
183 Read Digitaloutput None Byte 0 0 = Off
1 = Pulse
2 = Frequency
3 = Error level
4 = Error number
5 = Direction/Limit
6 = Batch
184 Write Relayoutput1 MAG 5000 C/6000 As command
Byte 0 0 = Off
3 = Error level
4 = Error number
5 = Direction/Limit
6 = Batch
7 = Cleaning
MASS 6000
Byte 0 0 = Off
3 = Error level
4 = Error number
5 = Direction/Limit