Gas consumption data
5E-E
Combustion gas Approx. 350 ml/min (approx. 21 l/h)
Hydrogen Approx. 20 ml/min (approx. 1.2 l/h)
Sample gas Approx. 1 l/min (approx. 60 l/h)
Zero gas/calibration gas Approx. 2 l/min (approx. 120 l/h)
5E-A (sample gas at standard pressure)
Combustion gas Approx. 350 ml/min (approx. 21 l/h)
Hydrogen/helium Approx. 110 ml/min (approx. 6.6 l/h)
Sample gas Approx. 1 l/min (approx. 60 l/h)
Zero gas/calibration gas Approx. 2 l/min (approx. 120 l/h)
5E-AS
Combustion gas Approx. 350 ml/min (approx. 21 l/h)
Hydrogen/helium Approx. 110 ml/min (approx. 6.6 l/h)
Sample gas Approx. 1 l/min (approx. 60 l/h)
Zero gas/calibration gas Approx. 2 l/min (approx. 120 l/h)
5E-I
Combustion air: None
Hydrogen Approx. 30 ml/min (approx. 1.8 l/h)
Sample gas Approx. 1 l/min (approx. 60 l/h)
Zero gas/calibration gas Approx. 2 l/min (approx. 120 l/h)
Oxygen cross-sensitivity 5E-E: Approx. 30 ppm C
1
Length Max. 30 m
Temperature 140 to 200 ˚C, freely-selectable
Power consumption (operation) Approx. 150 VA
Power consumption (startup) Approx. 350 VA