What does F5 error mean on Siemens Measuring Instruments and how to fix it?
- KKaren MartinezJul 26, 2025
Error code F5 indicates that the battery charge is below the preset limitation. Replace the battery pack to resolve this issue.
What does F5 error mean on Siemens Measuring Instruments and how to fix it?
Error code F5 indicates that the battery charge is below the preset limitation. Replace the battery pack to resolve this issue.
What does F3 error mean on Siemens Measuring Instruments?
Error code F3 indicates an internal software failure. Contact your supplier for assistance.
What does F4 error mean on Siemens Measuring Instruments?
Error code F4 indicates an internal software failure. Contact your supplier for assistance.
What does F1 error mean on Siemens Measuring Instruments?
Error code F1 indicates that there is no water in the upper part of the pipe and/or the cables or transducer 1A or 1B are defective.
What does F2 error mean on Siemens Measuring Instruments?
Error code F2 indicates that there is no water in the lower part of the pipe and/or the cables or transducers for 2A or 2B are defective.
What to do if Siemens Measuring Instruments display is blank?
If the display is blank, check the flowmeter version. Note that the battery version cannot run on mains power. Ensure the battery plug is connected, the battery is not empty, and the mains power is not interrupted. If you have a battery version, replace the battery pack.
Brand | Siemens |
---|---|
Model | SITRANS FUE380 |
Category | Measuring Instruments |
Language | English |
Essential safety guidelines for operating and maintaining the device.
Discusses responsibilities, design considerations, and safe installation practices.
States the device is not approved for hazardous area installations.
Outlines the two main steps for installing the flowmeter: sensor and transmitter.
Covers sensor installation, including environmental considerations and pipe conditions.
Specifies requirements for straight pipe sections before and after the flowmeter.
Describes how to install both compact and remote transmitter versions.
Details connecting a mains-powered transmitter, including backup battery options.
Step-by-step guide for connecting a compact flowmeter system.
Instructions for connecting the power supply cable for mains-powered units.
How to connect cables for pulse output signals.
Guide for connecting a remote flowmeter system.
Connecting the transmitter components in a remote system setup.
Explains the layout and basic operation of the device's local display.
Steps for the initial start-up of the device after power-up or battery replacement.
Detailed instructions on using SIMATIC PDM for device configuration and commissioning.
How to read and write device parameters using SIMATIC PDM for optimal settings.
Guidance on setting parameters like Pulse Output A and Qmax for system optimization.
Final steps to download parameters and verify device operation readiness.
Explains the use of a hardware key to access protected parameters.
Instructions and recommendations for replacing the device's battery pack.
Information on how to contact Siemens technical support for assistance.
Lists and explains error codes displayed by the device and in PDM.
How to use SIMATIC PDM for diagnosing device issues and analyzing data.
Specifies the typical accuracy of the SITRANS FUS380 and FUE380 flowmeters.
Provides detailed technical specifications for the FUS380 and FUE380 systems.
Technical specifications for the sensors used with FUS380 and FUE380.