EasyManua.ls Logo

Siemens GM-SG

Siemens GM-SG
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Circuit
Breaker
Positions
Racking
Crank
Engagement
Procedure
A
crank
for
racking
the
drawout
circuit
breaker
is
provided
as
a
standard
accessory
.
Racking
of
a
circuit
breaker
can
be
accomplished
with
the
circuit
breaker
compartment
front
door
open
or
through
a
small
opening
(
or
window
)
in
the
front
door
,
with
the
door
closed
.
The
racking
crank
consists
of
an
offset
handle
with
a
custom
socket
assembly
welded
to
the
opposite
end
.
The
socket
end
of
the
crank
is
designed
to
engage
the
square
shoulder
of
the
racking
mechanism
shaft
and
remain
engaged
during
racking
by
means
of
two
spring
plungers
located
180
degrees
from
each
other
.
The
spring
plungers
operate
in
a
manner
similar
to
the
retainers
on
an
ordinary
mechanic
'
s
socket
wrench
.
The
portion
of
the
racking
mechanism
shaft
which
is
visible
is
cylindrical
,
and
the
square
shoulder
of
the
racking
mechanism
shaft
is
hidden
by
a
shroud
until
the
engagement
procedure
starts
.
The
square
socket
end
of
the
crank
will
only
engage
the
square
portion
of
the
shaft
if
it
is
aligned
properly
.
The
suggested
procedure
to
engage
the
racking
mechanism
is
as
follows
:
1
.
The
circuit
breaker
must
be
open
.
(
The
racking
shroud
cannot
be
moved
if
the
circuit
breaker
is
closed
)
.
2
.
Hold
the
socket
end
of
the
crank
in
one
hand
and
the
crank
handle
in
the
other
hand
.
Refer
to
Figure
44
.
3
.
Place
the
socket
over
the
end
of
the
racking
mechanism
shaft
.
Align
the
spring
plunger
of
the
socket
with
the
shoulder
on
the
racking
mechanism
shaft
.
Note
:
If
the
socket
is
not
aligned
,
the
socket
will
not
be
able
to
engage
.
The
socket
(
square
)
must
be
aligned
with
the
square
shoulder
of
the
racking
mechanism
shaft
.
4
.
Once
alignment
is
achieved
,
firmly
push
the
crank
and
socket
assembly
toward
the
racking
mechanism
.
5
.
When
properly
engaged
,
the
crank
should
remain
connected
to
the
racking
mechanism
,
due
to
the
spring
plungers
.
If
the
crank
does
not
remain
in
position
,
adjust
the
spring
plungers
clockwise
one
-
half
turn
.
This
will
increase
the
contact
pressure
of
the
spring
plunger
.
6
.
To
remove
the
crank
,
simply
pull
the
assembly
off
of
the
racking
mechanism
shaft
.
Note
:
If
the
effort
to
rack
the
circuit
breaker
increases
considerably
during
racking
,
or
if
turning
of
the
racking
crank
requires
excessive
force
,
stop
racking
immediately
.
Do
not
try
to
"
force
"
the
racking
crank
to
rotate
,
or
parts
of
the
circuit
breaker
or
racking
mechanism
could
be
damaged
.
Determine
the
source
of
the
problem
and
correct
it
before
continuing
with
the
racking
.
Socket
Assembly
Hand
Position
Hand
Position
©
(
Spring
Plunger
Crank
Rotated
for
Clarity
Spring
Plunger
Racking
Mechanism
Shaft
-
rW
-
=
=
=
=
Articulated
Joint
Shoulder
Portion
of
Shaft
Custom
Socket
Spring
Plunger
Figure
44
.
Racking
Crank
Engagement
Procedure
.
Circuit
Breaker
Racking
When
inserting
a
circuit
breaker
into
a
cell
,
be
sure
that
the
racking
mechanism
is
in
the
disconnect
position
.
In
this
position
,
the
racking
position
indicator
should
show
a
green
square
with
the
letter
"
D
"
for
DISCONNECT
position
.
Important
:
Failure
to
follow
instructions
may
result
in
damage
to
equipment
.
Return
racking
mechanism
to
the
DISCONNECT
position
before
inserting
circuit
breaker
.
Figure
45
.
Racking
Mechanism
with
Racking
Block
.
42
Courtesy of NationalSwitchgear.com

Related product manuals