GENERAL ADVICE
12
The program values as preset from the factory are,
for Mo/W tube:
Program Compression kV-value Anode/Filter-
thickness combination
10 - 29 mm 26 Mo / Mo
2 30 - 44 mm 27 Mo / Mo
3 45 - 59 mm 27 Mo / Rh
4≈ 60 mm 26 W / Rh
for Mo tube:
Program Compression kV-value Anode/Filter-
thickness combination
10 - 29 mm 26 Mo / Mo
2 30 - 44 mm 27 Mo / Mo
3 45 - 59 mm 27 Mo / Rh
4≈ 60 mm 28 Mo / Rh
The program settings in the table are only suggestions. Each facility
may choose the most suitable setting for the film/screen in use.
The thickness intervals can only be changed by a service engineer.