Figure 4-5 Peak Screen - Calibration
As each peak is selected on the left pane, the calibration settings for that peak are displayed on
the right. The example below is designed to give the user a basic idea of modifying a calibration
and performing an offline calibration based on the modified settings.
Example Scenario:
A new calibration standard has been installed with slightly different concentrations than the
previous standard. You need to change the calibration levels and recalibrate the current
method.
User Actions:
Perform the following actions to change the calibration levels:
Select Peaks from the Method on the Navigation Menu and then select the first detector.
This displays peaks for the first detector.
Click the Calibration checkbox to see calibration settings.
Select the first peak (in this case Carbon Monoxide).
In the Calibration Level settings, change the Amount attribute to match the new calibration
standard.
Initial Startup Procedures
4.12 Calibrate the Maxum II
Installation Manual
46 Installation Manual, 12/2019, 2000595 Rev 1