MOBILETT Plus/ -E/ -HP SPR8-220.031.04 Page 12 of 30 Siemens AG
Rev. 04 09.03 CS PS 24 Medical Solutions
2 - 12 Installation
22. Splice the stripped cables in pairs of two with soldering coils according to the table
below and according to pos. 1 / Fig. 13.
23. Solder each of the six sets of spliced cables (pos. 2 / Fig. 13).
Twisted cable pairs: To cables from board D70:
White Brown
Purple Red
Twisted cable pairs: To cables from board D70:
White Black
Blue Blue
Twisted cable pairs: To cables from board D70:
White Yellow
Yellow Orange
Follow the figure above as closely as possible
.
Make sure the sol-
dering and stripped cables are completely covered by the shrink-
ing tubes to avoid short circuit.
MOB00609
Fig. 13 Soldering and shrinking procedure
21
3
4
NOTE