Installation 2 - 19
Siemens AG SPR8-220.031.04 Page 19 of 30 MOBILETT Plus/ -E/ -HP
Medical Solutions Rev. 04 09.03 CS PS 24
6. If needed, drill three holes with a diameter of 5 mm in the aluminum plate as
shown in Fig. 22 for the new power supply unit. The distances are: A = 65 mm, B =
46 mm, C = 32.5 mm, D = 108 mm and E = 25 mm. Also drill a hole (5 mm
diameter) as shown in pos. 1 / Fig. 22 to attach the cable to the new power supply.
7. Cut the cable ties as shown in Fig. 23.
If drilling is necessary:
Drilling chips can cause injuries to eyes.
Use proper protection.
CAUTION
MOB00521
A
B
C
D
Battery charger
Fig. 22 Adjustments for new power supply
E
1
MOB00522
Fig. 23 Cutting cable ties