Service Instructions
Multimobil 5C
Siemens Ltd. Med India Version 5.0
Page 30 of 73
Copyright © SIEMENS LTD. All rights reserved. For internal use only
Aspect ratio 4:3
Video output VS 1.0 V / 75 Ohm BNC connection, 2 routes
Lens mount C mount
Input power Single-phase, 100 - 240V, 50 or 60Hz
Power consumption 30 W approx.
Environmental
Temperature Humidity Air pressure
Operation
+10 to +40 °C
30-75% RH max. 70 – 106 kPa (no
condensation)
Storage &
Transportation
-15 to 45 °C
95% RH max. 76 – 106 kPa (no
condensation)
6.3 CONNECTION TERMINALS OF CAMERA CONTROL UNIT (CCU)
CAMERA (D-sub15) Connected to the camera head with the camera
cable.
VIDEO OUT 1 and 2 (BNC) Digital Video signal: LIH image, Memory image.
Monitor is connected to this terminal.
REMOTE CONTROL 1 [36-
pin(D57;DDK)]
Input/output of the one signal group for external
control of camera.
REMOTE CONTROL 2 (D-sub25) Input/output of the other signal group for
external control of camera.
HD OUT (BNC) Horizontal drive pulse output for external
devices
VD OUT (BNC) Vertical drive pulse output for external devices
AC100-240V AC power input.
FG Grounding terminal.
CONT (D-sub9) For camera control with RS232c.
6.4 PERFORMANCES /FUNCTIONS
CCD operation method Frame accumulation
Resolution (CCIR) 560 TV lines (H) 520 lines (V) or more
for frame accumulation
S/N ratio
55dB or more (under Normal gain, AP;OFF, γ
;1, Using 4.2 MHz low pass filter)
Frequency response 0 to 8 MHz (-3dB)
Aperture correct. (AP) Front & rear edge emphasis for both H&V. AP
ON/OFF & emphasis level are selectable