Service Instructions
Multimobil 5C
Siemens Ltd. Med India Version 5.0
Page 70 of 73
Copyright © SIEMENS LTD. All rights reserved. For internal use only
8.4.9 CodE 22: Maximum Preparation Time
START
TRIGGER A RADIOGRAPHY
IS CODE 22
DISPLAYED?
RESET THE ERROR
MONITOR THE J36.15 SIGNAL
IS SIGNAL LEVEL
CHANGES/
CONFIRM THE SWITCH WIRING OR
CHANGE HAND HELD EXPOSURE
SWITCH.
STOP
Yes
No
Yes
No
CHANGE D915