8.3 Standby Error codes
8.3.1 CodE 02 : +15 V Supply Error
START
REMOVE CONNECOR X15 ON D915
& SWITCH ON THE UNIT
MEASURE +15 V ON X15
CONNECTOR
IS +15V THERE ?
YES
REPLACE D915 PCB
NO
REPLACE SMPS1