Area dose product measurement device 3
Siemens AG SPR2-240.814.04 Page 3 of 10 POWERMOBIL/ARCOSKOP
Medical Engineering Rev. 02 07.00 TD SD 24
Installation 1
Disconnect the POWERMOBIL respectively ARCOSKOP from the mains power.
Remove the POWERPHOS cover.
Without the laser targeting device: 1
Attach the Area dose product measurement device to the POWERPHOS using the
screws included in the packaging.
Connect the interface cable already integrated in the cable harness to the Area dose
product measurement device and lock it in place.
Reattach the POWERPHOS cover.
With the laser targeting device: 1
Remove the laser targeting device from the POWERPHOS by removing the screws.
Mount the Area dose product measurement device to the POWERPHOS.
Reattach the laser targeting device.
Connect the interface cable already integrated in the cable harness to the Area dose
product measurement device and lock it in place.
Check the laser targeting device according to the adjustment instructions and,
if necessary, adjust it.
Reattach the POWERPHOS cover.