POWERMOBIL/ARCOSKOP SPR2-240.814.04 Page 8 of 10 Siemens AG
Rev. 02 07.00 TD SD 24 Medical Engineering
8 Area dose product measurement device
Evaluation 1
Calculate the deviation between the measured area dose product and the displayed area
dose product:
The deviation must be < 0.3.
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg)] /
(measured area dose product FDPg).
Fig. 3 Fig. 4
Fig. 5
Measuring chamber
POWERPHOS
RBV / I.I.
Centering cross
POWERPHOS
RBV / I.I.