27
To set segment 6 (°C)
Temperature from 55°C to 30°C
Duration 2h - slope - gradient 0.21°C/min
hh : mm : ss
Durat. 2 : 00 : 0
Wait duration : ON
G1
Channel 1
Control : ON
Wait set point : ON
Max speed : OFF
Set point : 30.0 0 °C
Grad. : 0.21 °C/min
Near set : 1,00 °C
G2
G3
Because the same temperature value
is maintained
OP2605-090
Channel 2
Control : ON
Wait set point : OFF
Max speed : ON
G4
G5
OP2605-100
Set point : 95.00 RH %
Grad. : 20,0 RH %/min
Near set : 3,00 RH %
G8
to save the programmed segment
Save segment : SAVE
G9
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 7
Wait for this number to appear
Water recharge: OFF
Not used: OFF
Not used: OFF
SP. hum. CTRL. ON
To improve the system performance
(experimentally set)
G6
To set segment
6 (% RH)
Humidity from 90% to 95% and maintenance
Duration 2h - Maintenance
6
Unreal gradient which enables maxi-
mum speed to be obtained
Aux 1: OFF
Aux 2: OFF
Aux 3: OFF
Aux 4: OFF
In this segment the AUX1 must be
switched OFF
G8
SHIFT
ENTER
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
This page is visualized.
Save segment : OFF
Total segment : 0
Confirm progr : OFF
When this page is visualized, if the previ-
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
Move on to the next step.
6