16
10
15
20
25
30
35
5
5’
10’
15’
t= min.
Temp
°C
3.2 STEP C Set a controlled humidity slope from 60% to 90% with a gradient of 2%/min with a
constant temperature and 60% RH humidity
Wait for the chamber to reach the preset values, 35°C
and 60% RH (Step A and Step B).
Follow the example given to programme a temperature
slope as shown in the diagram with the gradient indi-
cated.
Move to the next page.
Temperature slope
already carried out
K2
Settings
A1
Temp. Set 35.0°C
Humid. Set 60.0 %
A2
The previously set values are visual-
ized.
Temp. gr. 0.00°C min
Hum. gr. 2.00 RH%min
Turn the temperature gra-
dient to 0
Turn the humidity gradi-
ent to 2
A4
ENTER
A3
Temp. Set 35.0°C
Humid. Set 90.0 %
ENTER
If the K1 key lights up, the machine will
carry out the command immediately
Final humidity
40
50
60
70
80
90
30
15’
20’ 25’
t= min.
% RH
KEYS
DISPLAY FUNCTIONS
gradient:
2%/min
30’