14
MAN-100411
SITRANS LVL200E - Operating Instructions
55328-EN-170712
1
Fig.5:Forkpositionwithangeversions
1 Markingwithangeversion,facingup
Use the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cableentry.Rainandcondensationwatercanthusdraino.Thisap-
plies mainly to outdoor mounting as well as installation in areas where
high humidity is expected (e.g. through cleaning processes) or on
cooled or heated vessels.
To maintain the housing protection, make sure that the housing lid is
closed during operation and locked, if necessary.
Makesurethatthedegreeofcontaminationspeciedinchapter
"Technical data" meets the existing ambient conditions.
Fig. 6: Measures against moisture ingress
Caution:
Do not hold SITRANS LVL200E on the tuning fork. Particularly with
angeortubeversions,thetuningforkcanbedamagedjustbythe
weight of the instrument. Transport coated instruments very carefully
and avoid touching the tuning fork.
Remove the packaging or the protective cover just before installation.
Moisture
Transport