EasyManua.ls Logo

Siemens SITRANS LVL200E - Page 28

Siemens SITRANS LVL200E
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
MAN-100411
SITRANS LVL200E - Operating Instructions
55328-EN-170712
3
2
1
16 ±1,5
I/mA
8
±1,5
< 3,6
0,6
±0,2 1,5 ±0,51,5 ±0,5
t/s
Fig. 29: Flow chart of the function test
1 Fault message
2 Empty signal
3 Full signal
Check if all three switching conditions occur in the correct sequence
and the stated time period. If this is not the case, there is a fault in the
measuring system (see also the operating instructions manual of the
signal conditioning instrument). Keep in mind that connected instru-
ments are activated during the function test. By doing this, you can
check the correct function of the measuring system.
Note:
Keep in mind that the starting time t
A
of the voltage supply can extend
thetimeuptotherstswitching(e.g.SITRANSSCSC,SITRANS
TCSC +1 s)
Test procedure
After releasing the button or after a brief line break.
Sensor
current -
Sensor
Level relay
A-overll
protection
Signal lamp
A-Overll
protection
Level relay
B - dry run
protection
Signal lamp
B - Dry run
protection
Fail safe
relay
Control
lamp
1. Fault sig-
nal
0.6 s
(±0.2 s)
+ t
A
1)
< 3.6 mA currentless currentless currentless
2. Empty
signal
1.5 s
(±0.5 s)
8 mA
(±1.5 mA)
energized currentless energized
3. Full signal
1.5 s
(±0.5 s)
16 mA
(±1.5 mA)
currentless energized energized
1)
Starting time of the voltage supply

Other manuals for Siemens SITRANS LVL200E

Related product manuals