Result Range: 1.1 - 33.3 mmol/L
Sample Size: 0.5 microliter (0.5 μL)
Sample: Fresh capillary whole blood or control solution
Test Time: Results in as fast as 4 seconds
Result Value: Plasma values
Assay Method: Electrochemical
Power Supply: One 3V lithium battery #CR2032 (non-re-
chargeable), Maximum rated current: 10mA
Battery Life: Approximately 1000 tests or 1 year
Manufacture date: Refer to meter label.
Shelf life: 10 years (based on every day perform 7 tests)
Software release version: V01
Memory Size: 500 glucose results
Size: 40mm×36mm×22mm
Operating Range
Relative Humidity: 10%-80% (Non-condensing)
Temperature: 5°C - 40°C
Hematocrit: 20%-70%
Storage condition: 10~93%RH, -20~55℃
Altitude: Up to and including 3108 meters
11. Specification
24
Table 2:Repeatability Requirement
1.10mmol/L~5.55mmol/L
(20mg/dL~100mg/dL)
SD<0.34mmol/L(6 mg/dL)
PrecisionTest Range
CV<6%
5.55mmol/L~33.30mmol/L
(100mg/dL~600mg/dL)