24
AUTOMATIC SETUP
1. Ensure that the process is in normal conditions.
2. Make sure that the monitor has been powered for at least 15 minutes in
order to warm up and stabilise.
3. Go to service mode if S305. S304 service mode is not needed
4. Press the AUTOMATIC SETUP BUTTON (the white button on lower
printed circuit board)
5. Make sure that green LED indicator turns red
6. Mount the cover of enclosure and tighten it and wait for about 45
minutes until the LED turns green indicating that the Automatic Setup
sequence has finished
7. The unit is ready to use
6.7 The manual range (MR) setup (Only S304)
The manual range is only for models 304. The manual range parameter set up
is found in parameter table in 304 models just before saving the parameters (-
SP). The MR parameter value is displayed in left digit as 1.. The MR
parameter is not in use as factory setup, but it can be activated by using other
than 00 value and by saving the change with āSP command. The MR setup is
used if the process is not stabile and the automatic set up can not be made or
if the auto setup 100% range is not big enough to monitor the dust
concentration range in the process.