6702 HERCULES Owner's • REV6
2-1. Intended Use
This surgical table is intended for use by healthcare
professionals for human medical purposes only.
The surgical table is not intended to be used for
patient transport.
2-2. Installation
Prior to placing the table into use, the following
items must be inspected, veried, and calibrated
by an authorized Skytron representative:
Final initialization and completion of the
installation report is required for warranty
validation.
Functional testing and cycling
Electrical safety testing to include
verication by hospital personnel
Digital calibration of the hydraulic system's
pressure relief valve (PRV)
Inspection of the hydraulic system
Table must be allowed to acclimate to
usage climate requirements
Verication of hydraulic uid level
Table has been wiped down to remove
rust inhibitor
Items found to be non-conforming must be
addressed prior to placing the table into service.
2-3. Environmental Conditions
a. During Transport and Storage (In Original
Packaging Materials)
Ambient Temperature: 14° to 122°F (-10°
to 50°C)
Relative Humidity: 10% to 85% (No
Condensation)
Atmospheric Pressure: 21 in-Hg to 31
in-Hg (700 hPa - 1060 hPa)
During Use
Ambient Temperature: 50° to 104°F (10°
to 40°C)
Relative Humidity: 30% to 75% (No
Condensation)
Atmospheric Pressure: 21 in-Hg to 30
in-Hg (700 hPa to 1000 hPa)
NOTICE
Operating altitude is 6562 feet (2000
m) max. above sea level.
2-4. Certication
Certied by ETL to these standards:
Medical electrical equipment, Part 1: General
requirements for basic safety and essential
performance ANSI/AAMI ES60601-1:2005 +
C1:2009 + A2:2010 /(R)2012
Medical electrical equipment–Part 1: General
requirements for basic safety and essential
performance CAN/CSA-C22.2 No. 60601-1:08 +
COR 2: 2011/06/01
Medical electrical equipment Part 2-46: Particular
requirements for the basic safety and essential
performance of operating tables IEC 60601-2-46:
2010
2-5. Classication
Class I Equipment
Applied Parts: Table Top/Type B Applied Parts
IPX4 Rated
Equipment not suitable for use in the
presence of ammable anesthetic mixture with
AIR, OXYGEN, or NITROUS OXIDE.
This product is not intended for sterilization.
2-6. Electrical Specications
Power Requirements: 100 - 240 VAC, 50 - 60 Hz,
400 VA
Current Leakage: Less than 500 micro amps
Power Cord: 15 foot (4.5 m) w/ hospital grade
connector (removable)
Duty Cycle: 3 min on, 7 min o󰀨
Battery Power: 24 VDC (12Vx2) Model : FML12170-
12V17Ah/Furukawa Battery OR PS-12180 F2
12V18Amp.Hr/PowerSonic (SKYTRON Part
Number : E0002293)
2-7. Mechanical Specications
Maximum Lifting Capacity: 1200 lbs (540 kg)
Maximum Articulating Capacity: 1000 lbs (450 kg)
Unit Weight: 750 lbs (340 kg)
Maximum Patient Weight: (1200 lbs (540 kg)
SECTION 2. EQUIPMENT SPECIFICATIONS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Skytron 6702 HERCULES and is the answer not in the manual?

Skytron 6702 HERCULES Specifications

General IconGeneral
BrandSkytron
Model6702 HERCULES
CategoryMedical Equipment
LanguageEnglish

Related product manuals