Page 30
6702 HERCULES Owner's • REV6
4-1. Cleaning and Disinfecting
WARNING
Personal injury to patient or sta󰀨
may result from a lack of proper
maintenance of this equipment.
CAUTION
Caution should be taken when cleaning
the table to prevent excessive uid entry
into electrical connectors.
NOTICE
Always follow current AORN/EORNA
Journal Guidelines to ensure proper
cleaning and disinfection procedure.
Always follow product cleaning and disinfecting
instructions and warnings provided by the
cleaning product manufacturer.
a. Cleaning
The following procedure should be followed when
cleaning the surgical table between cases or operations:
1. Place table top in the level position.
2. Ensure that all power is removed from the table.
Take preventive measures to avoid spraying directly
into connectors or electrical receptacles.
WARNING
Always follow OSHA/EASHW blood-
borne pathogens standards for
protective clothing, including gloves,
masks, and eye protection when
cleaning the surgical table.
3. Remove major contaminants from the table
with disposable materials following appropriate
biohazard waste disposal procedures.
4. Remove all table pads and place them on
a at surface for cleaning.
CAUTION
Thoroughly read and follow manufacturer's
directions for all cleaning uids. DO NOT
use cleaners containing phenolics.
5. Use a ready-to-use detergent diluted as
required by manufacturer. Ensure that the active
ingredients of the product are compatible with
the materials of the SKYTRON table.
6. Apply cleaning uid liberally to top and sides
of each pad and wipe with a clean lint-free
cloth.
7. Using a clean, damp, lint-free cloth, wipe
the pads to remove the cleaning uid.
8. Using a clean, dry, lint-free cloth, wipe the
pads to remove all moisture.
9. Repeat steps 6 through 8 to clean the bottom
of each pad. Allow to dry.
CAUTION
When using spray cleaners, DO NOT
spray fluids directly into electrical
receptacles or components.
10. Repeat cleaning procedure for all table
surfaces including the top, sides, elevation
column, base, and all accessories.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
11. When the cleaning procedure is complete, replace
all pads and accessories as applicable.
CAUTION
Avoid immersing the pendant control
assembly in liquids.
12. Apply cleaning solution to the pendant
control and cord.
13. Use a clean cloth dampened with water to
remove cleaning solution.
14. Use another clean damp cloth to remove
any remaining residue.
15. Install pendant control on side rail for storage
when cleaning procedure is complete. Allow
to dry.
16. Clean casters and oor lock brakes.
SECTION 4. MAINTENANCE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Skytron 6702 HERCULES and is the answer not in the manual?

Skytron 6702 HERCULES Specifications

General IconGeneral
BrandSkytron
Model6702 HERCULES
CategoryMedical Equipment
LanguageEnglish

Related product manuals