5
Product description, cont’d
1.3 Technical data, PFM FLEX Concept.
Measurement Range
- Total Pressure: max 2500kPa
- Differential Pressure: 0-1000kPa. Optional: 0 - 2000kPa
- Recommended pressure range: 3 – 200kPa
- Static pressure: <1000kPa Optional: <2000kPa
- Ambient temp: -30 - +40°C
- System Temp T1: -30 - +120°C
- System Temp T2: -30 - +120°C
Measurement deviation
- Differential Pressure: <0,1% of Full Scale
- Flow: as for differential pressure + valve deviation
- Temperature: <0,2°C
Battery capacity, operating and charge time
Hand Terminal
- Battery Capacity: 1200mAh. Extended: 3900mAh. Optional.
- Operating Time: 8 hours. Extended: 24 hours, continuous duty.
- Charging time: 1,4 hours. Extended: 4,5 hours
Sensor:
- Battery Capacity: 6.600mAh
- Operating Time: 35 hours, continuous duty
- Charging Time: 6 hours
Ambient temperature for the instrument
- During operation and charging: 0 – 40°C
- During storage: -20°C – 60°C
Humidity Ambient humidity: max 90%RH
Weight Sensor: 540gr Complete Case: 2,8kg
Open Space Range: >200m. >500m with antenna (option)
IP-class Sensor: IP65