iGlucose® 33
study results of typical iGlucose users
Like all commercially available blood glucose meters, your
iGlucose® Blood Glucose Meter result may vary slightly from
your actual blood glucose value. This may be due to slight
dierences in technique and the natural variation in the test
technology. The table below shows the results of a study
where 153 typical users used the iGlucose Meter to test their
blood glucose level. When glucose results were below 75 mg/
dL the iGlucose Meter gave results within 15 mg/dL of their
true blood glucose level 6 out of 6 times.
Dierence range between
the true blood glucose level
and the iGlucose Meter result
Dierence range between
the true blood glucose level
and the iGlucose Meter result
The percent (and number) of
meter results that match true
blood glucose level within x%
The percent (and number) of
meter results that match true
blood glucose level within x%
When the users’ results were equal or greater than 75 mg/dL in
this study, the iGlucose Meter gave results within 15% of their true
blood glucose level 140 out of 147 times.
33.3%
2/6
43.5%
64/147
Within ±5
mg/dL
Within ±5
mg/dL
83.3%
5/6
81.6%
120/147
Within ±10
mg/dL
Within ±10
mg/dL
Within ±15
mg/dL
Within ±15
mg/dL
100%
6/6
95.2%
140/147