22 • ENGLISH
Problem Possible Cause Solution
Cuff is too tight or not
properly positioned on
the wrist.
Inaccurate test results due to
body movement or monitor
movement.
Cuff fails to inate properly.
Improper operation.
Pressurization is over
cuff rated pressure
300mmHg
Blood pressure
results are not
within typical
range
“Err” displayed
Adjust the cuff 1/4” - 1/2”
from the edge of
the head of the ulna bone.
Sit in a relaxed position with
arm placed near heart. Avoid
speaking or moving body parts
while testing. Make sure the
monitor unit is placed in a
stationary position throughout
the testing period.
Make sure hose is properly
fastened to cuff and monitor
unit.
Read user manual carefully and
re-test properly.
Read user manual carefully and
re-test properly.
TROUBLE SHOOTING