ENGLISH • 25
ELECTROMAGNETIC COMPATIBILITY
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specied below.
The customer or the user of this monitor should assure that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
monitor requires continued operation
during power mains interruptions, it
is recommended that this monitor be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment
Guidance