Applicable Fluids and Precautions
Applicable fluid list
Substance description
Judgment
Remarks
Water
O
Conductivity of tap water: 100 to 200μS/cm
Deionized water
X
The electric conductivity is too low.
Water base coolant
O
When the ratio of water is 50% or more.
Oil
X
The electric conductivity is too low.
Oil base coolant
X
The electric conductivity is too low.
Sea water
X
Corrosive to the product.
Ethylene glycol
X
The electric conductivity is too low.
Ethanol
X
The electric conductivity is too low.
Methanol
X
The electric conductivity is too low.
Chloride water
(Hypochlorous acid)
X
Corrosive to the product.
*: The table is for reference only. O: Acceptable X: Not acceptable
(1) Operate fluids with electric conductivity of 5μS/cm or more.
The electric conductivity is a ratio which shows how easily the electricity flows.
Note that this product can not be used for fluids with low conductivity.
This product cannot be used for fluids that do not conduct electricity such as De-ionized water and oil.
(2) If insulating material gets stuck inside of the piping, it may cause an error.
Remove the foreign material stuck inside of the piping with a brush for washing test tubes so that internal resin piping
will not be damaged.
(3) If conductive materials such as metal get stuck to the interior surface in the piping, the switch may
malfunction.
Remove the foreign material as mentioned above.
(4) If stray electrical currents are flowing through the fluid to be measured, the switch may malfunction.
Beware that electrical leakage currents may be generated by equipment around the flow sensor such as pumps,
valves and metal piping when this equipment is at different electrical potentials in relation to earth ground.
(5) Any fluid which corrodes the internal fluid contact parts cannot be used.