-7-
No.LFE****-OMZ0012
- Design and install the product so that fluid always
fills the detection passage.
1. If the detection passage does not become fully
filled with fluid during the use of the product, an
incorrect detection signal can be generated in the
electrodes, prohibiting correct measurement. Be
sure to install the product so that fluid will remain
in the detection passage even after the fluid flow is
stopped.
For vertical mounting, apply fluid from the bottom
to the top. Bubbles may be generated when
applying fluid from the top to the bottom, leading
to operation failure.
(There should not be a problem as long as the
fluid passage is completely filled with fluid)
2. When the product is mounted horizontally, place the cover perpendicular to the floor (to place the electrodes
on the right and the left) to unaffect bubbles.
Fluid passage
Susceptible to bubbles
Not susceptible to bubbles
Above
Down
Down
Cover
Cover
Above
Bubbles
Bubbles
Electrode
Electrode