25
Troubleshooting
Note: If any of following issues persist, return wrench to an authorized Snap-on repair center.
Wrench does not turn on when
POWER button pressed.
Cycle power using end-cap
Torque reading out of spec.
Incorrect head length
entered
Enter correct offset head
length
Wrench did not retain settings
while batteries were removed.
Batteries removed before
setting were saved in non-
volatile memory.
Clear data, re-enter settings
and press and hold POWER
button to power down
wrench before removing
batteries.
Press ENTER button to
continue using wrench and
replace batteries soon.
Press POWER button to
turn off wrench and replace
batteries.
Torque applied while zeroing
Remove torque and re-zero
Wrench improperly
calibrated
Wrench moving during
zeroing
Place wrench on stable
surface
ENTER button pressed during
angle zeroing (Aborted
zeroing to access menus)
Press POWER button to re-
zero
Over 125% of full scale
torque applied
Cycle power using POWER
button and recalibrate
Wrench rotated too fast
during angle measurement
Press POWER button to re-
zero
Calibration interval exceeded
or invalid date entered with
calibration interval enabled
Calibrate wrench or press
ENTER to continue. Disable
calibration interval if not
required.