74
English
BCD
Alarm terminal is
“H”
True logic: fifth
digits are all “H”.
False logic: fifth
digits are all “L”.
Alarm terminal is
“H”
RS-232C
∗E
C
R
L
F output.
(∗ is A or B.)
∗MFX.XXX
C
R
L
F output.
(∗ is A or B, M is a space
or a minus sign.)
∗E
C
R
L
F output.
(∗ is A or B.)
Output
9. Alarm display/output
Terminals
All “H”
—
All “H”
Item
Measuring probe
is not connected or
connection has
been cut.
Acceleration over
Signal disorder or
connection cut
Overflow
Initialization is
required.
LCD display
Fifth digit is F
Value for axis is flashing.
Solution
Reset
Fix problem and
reset.
Reset
Fix problem and
reset.
Limit input to 5 digits
and reset.
Move the measuring
unit at least 1 mm
and reset.
Cause
The measuring probe was
exchanged with the power on.
Measuring probe is not
connected or connection
has been cut
The measuring probe’s
spindle exceeded the
maximum acceleration.
Abnormality in measuring
probe’s output signal or
connection has been cut.
Value exceeded 5 digits.
Initialization is required to
calibrate the new pairing of
both device and measuring
probe.
Note
If the alarm is again displayed/output after solving the problem
Has the measuring probe or its spindle received some strong shock?
Try replacing with the measuring probe used for the other channel.